The rotameter is also called a float flowmeter. It is a variable-area flowmeter. The liquid flows in the scale tube of the flowmeter. From bottom to top, the buoyancy of the fluid and the upward lift drive the float to rise, but the gravity reaches At a time of balance, the float remains stable, and the flow indicating scale at this time is the actual flow of the flow.
However, there are some faults in the rotameter sometimes, and there are three common types:
First, the rotor flowmeter measurement error is large:
1. Since the gas medium is greatly affected by temperature and pressure, it is recommended to use temperature and pressure compensation to obtain the true flow rate.
2. Due to long-term use and pipeline vibration and other factors, the floating magnetic flowmeter sensing magnetic steel, pointer, counterweight, rotating magnetic steel and other moving parts are loose, resulting in large errors. Solution: You can verify by hand pushing the pointer first. First press the pointer to the RP position to see if the output is 4 mA, whether the flow display is 0%, and then verify according to the scale. If the discrepancies are found, the parts can be adjusted. Generally require professional adjustment, otherwise it will cause the position to be lost, need to return to the manufacturer for correction.
3. Installation does not meet the requirements: for the vertical installation of the rotameter to maintain vertical, the inclination angle is not more than 20 degrees; for horizontal installation of the rotameter to maintain horizontal, the inclination angle is not more than 20 degrees; there must be no ferromagnetic objects in the 100mm space around the rotameter . Install the position away from the valve reducer, pump outlet, process line turn, etc. To maintain the requirements of the 250mm straight pipe section after the first 5D.
4. The large change in density of the liquid medium is also a cause of large errors. Since the instrument is converted according to the density given by the user before the calibration, the flow rate of the water in the calibration state is calibrated. Therefore, if the density of the medium changes greatly, the measurement will cause a large error. The solution can bring the density of the medium after the change into the formula, convert it into the error correction coefficient, and then multiply the flow measured by the flow meter by the coefficient to replace the actual flow.
Second, the rotor flowmeter pointer jitter:
1. Slight pointer jitter: generally caused by media fluctuations. It can be overcome by increasing the damping.
2. Moderate pointer jitter: generally due to media flow conditions. Gases are generally caused by unstable media operating pressures. A steady or steady flow device can be used to overcome or increase the rotor meter air damping.
3. Violent pointer jitter: mainly due to medium pulsation, unstable air pressure or user-supplied gas operating state pressure, temperature, flow rate and the actual state of the rotor flowmeter do not match, there is a large difference caused by the rotor flowmeter over-range.
Third, the rotor flow meter pointer stops at a certain position:
The main reason is that the rotor of the rotameter is stuck. Generally, when the rotameter is used, the valve is opened too fast, so that the rotor quickly impacts the stopper upward, causing the stopper to deform and the rotor to be stuck. However, it is not excluded that the rotor is stuck due to the difference between the rotor guide rod and the stop ring. The instrument can be removed during processing, the deformed stopper is removed and shaped, and checked whether it is concentric with the guide rod. If the heart is not correct, the rotor can be installed and the rotor can be pushed by hand to feel the smooth and unobstructed card of the rotor. Yes, in addition, when the rotameter is installed, it must be installed vertically or horizontally, and it cannot be tilted. Otherwise, it may easily cause the card table and bring errors to the measurement.
EVA Embossed Sheet is also called micro dot sheet, made by a roller heat press for the surface texture. It has same function as the EVA Faux Teak Sheet, providing a non-skid feature and perfect protection for boat.
Obviously it is also a DIY marine sheet for boat lovers, cut the sheets into size to fit the shape of boat decking, then just simply peel and stick. The micro dots on the surface not only provides a soft and comfortable feeling to feet, but also brings a suitable traction. Due to its sealed embossed surface, it is a lot easier to clean than the EVA faux teak sheet with brushed surface.
Eva Embossed Sheet,Coaming Bolster Pads,Embossed Eva Foam Sheet,Eva Coaming Pads,Embossing Eva Rubber Sheets,Eva Boat Flooring Sheet
Huizhou City Melors Plastic Products Co., Limited , https://www.thefoammaster.com